MODEL 2032 HIGH ACCURACY AUTO RANGING SCANNING DVM

FEATURES
* High Accuracy - 14 Bits of Resolution with Microprocessor Controlled Ditherized
* Auto Ranging Amplifier Provides Dynamic Range of 1 uV Resolution vs 10 V Full Scale
* Auto Calibration via Internal Voltage Reference and Microprocessor
* 32 Differential Channels in a Single Width Module
* Good 50/60 Hz Rejection - Programmable Digital Filter
* Self Diagnostics

DESCRIPTION

The Model 2032 is a complete voltage measuring instrument. With a resolution down to 1 uV, a programmable digital filter for 50/60 Hz noise, and an autoranging amplifier, the 2032 provides for the measurement of 32 differential inputs in a single width CAMAC module.

The onboard microprocessor averages 64 measurements in 1/60 (1/50) seconds to suppress power line noise. Each measurement uses a ditherizing technique which adds to the signal, then subtracts from the result, a random offset to average out the differential non-linearity of the ADC. Self-diagnostics and automatic calibration of gain and offset are also performed by the microprocessor.

Front-end voltage protection and filtering, common-mode rejection and additional normal mode rms measurement capability add to the usefulness of this module in high noise environments for monitoring power supplies, for strain gauges, thermocouples, flow gauge measurements, and a wide range of other applications.

TECHNICAL SPECIFICATIONS - MODEL 2032

ANALOG INPUT
 
Channels 32 differential inputs
Impedance: 20 MOhm.
Signal Conditioning 200 Hz roll off filters in each channel. Automatic ranging differential amplifier with gains from 1 to 1024.
Common Mode Rejection:  80 dB
Common Mode Range:  10 V for specified accuracy
Overvoltage Protection: 1 00 V for 1 sec.
CONVERSION
 
Analog-to-Digital Conversion: 25 uSec
Basic Resolution: 12 bits (1 part in 4096). Dithered resolution after averaging = 14 bits
Scanning Mode: Each channel is updated every 640 mSec. 64 measurements/16.7 mSec in each channel.
DATA MANIPULATION
 
Digital Filter: 64 successive samples are averaged together over a 16.7 mSec interval to suppress 60 Hz variations.
Note: 50 Hz version available as option.
Auto-Ranging: Gain of input amplifier is microprocessor controlled to autoscale digitized voltage between 40% and 90% of full scale.
Auto-Calibration: Built-in voltage reference is used to continuously monitor gain and offset which is factored into the measurements, (once per 2 minutes).
Ditherizing: Random value of a small voltage is superimposed on the input signal by a DAC. The digital value of the DAC voltage is then subtracted from the digitized ADC value.

COMPUTER COMMANDS
X is returned for valid CAMAC commands.
X = 0 if calibration  or  AC  measurement  is  in progress.
F (0): Read Output of  channel  n  set  by  F(17) and  increment  channel  number.   Q=0   if   data not available.  Q  Scan  feature:  Q=0  following read of channel 32.
F(9): Processor reset
F(16): Write  Control  Register  to  change  scan rate, read software  revision  number  or  select IEEE or VAX output data format.
F(17): Sets first channel  to  be  read.  Channel number is placed on data lines (0-31).

MATING CONNECTOR
2  ea.  Amphenol  204731-2   (P/N   MC3016)   are included.

PACKAGING:
#1 width CAMAC module
221 mm H, 18 mm W, 292 mm D* (8.7'x 0.7" x  11.5"
*Depth  front  to  rear  panel.  Rear   connector 13 mm (0.5").
In  conformance  with  the  CAMAC  standard   for RF   shielded   instrumentation   modules   (IEEE standard    585,    European     Esone     Report #EURA\4100e.)

POWER REQUIREMENTS:
       + 6V       900 mA
       +/- 24 V     50 mA

TEMPERATURE RANGE
0' C to 400 C  (32'  F  to  1040  F)  ambient  to operate within specifications when  installed  in crate  with  enough  air  flow  to  hold  maximum air exit temperature to 550 C (131' F).