MODEL 1880B DUAL READOUT SCALER
The Model 1880 is a dual NIM width module containing two identical
channels of eight (8) digit BCD 100MHz scalers. Readout is provided
on eight (8) in line LED readouts for each channel.
SPECIFICATIONS
CAPACITY Eight (8) digits BCD plus overflow.
SIGNAL INPUT(each channel)
Pulse Repetition Rate: 0 to 100mpps min(125mpps typical)
Pulse Pair Resolution: 10 ns min.
Minimum Pulse Width: 3nsec FWHM
Discriminator Level: -500mv (internally adjustable to a
min. -300mv)
Input Impedance: 50Ohm 10% D.C. coupled
Input Reflections: Less than 10% for lns rise time of
NIM level signals.
GATE (each channel)
Manual: Front panel three position switch
(OPEN,REMOTE,CLOSED) for controlling
the counting gate. In the remote
position electrical gate signals
control the counting gate.
Electrical: Two bridged LEMO connectors on front
panel. Bridged inputs allow daisy
chaining of gate signals from scaler
to scaler.
Input Impedance: 5K min.
Sensitivity: -500mv
Input Protection: t 50 volts transients
Response Time: Less than 10nsec.
RESET (each channel)
Manual: Front panel push button for reset of
counter overflow and display.
Electrical: Two bridged LEMO connectors on front
panel. Bridged inputs allow daisy
chaining of gate signals from scaler
to scaler. The input is biased to a
CAMAC logic zero (approx. 3.5v) and
a ground or a current sink of 10ma
will reset scaler channels to zero.
DISPLAYS (each channel)
Counter: Eight (8) inline 3/8" high character
LED seven(7) segment digits.
Gate: LED light on when counter gate is
open.
Overflow: LED light on when counter overflows.
Turns off when counter is reset.
OUTPUTS (each channel)
Carry Output: LEMO front panel connector provides
a fast NIM pulse -800mv into 50 ohms
each time counter overflows. A
carry pulse will not be generated
at reset.
POWER REQUIREMENTS: Less than 12 watts
+ and - 6 volts
TEMPERATURE RANGE 0 to 60'C
SIZE Double width 8 3/4 high NIM module.
|